Urology Instruments
What is the most delicate part of the stone crushing forceps.
Shaft
Distal Tip
Jaw
Hinge
Minor damage may be repairable. Major damage will require the instrument to be replaced.
Can dents be repaired on the shaft of optical biopsy forceps.
Yes
No
The distal tip, where the ingle-action jaw is connected to the shaft. The area around the securing pin must be inspected for cracks on both sides prior to and after every use.
What causes blurry image on a cystoscope.
Cleaning
Drying
Autoclave
Transport
Cystoscopes have a series of glass rod lenses that are commonly damaged during use and transport. If one of the lenses gets cracked, a blurry image will result. The repair vendor can replace the cracked rod lens.
Can the ceramic tips be replaced on Resectoscope sheaths.
Yes
No
The distal tip of a resectoscope sheath must be inspected prior to and after each use for cracks and chips. If damage is found, the instrument must be immediately removed from service and sent out for repair.
Which instrument is fitted to sheath for inserting the scope and forceps.
Deflecting Obturator
Stone Crushing Forceps
Double Channel Bridge
Cystoscopy Biopsy Forceps